User Manual
Table Of Contents
- Revision History
- Table of Contents
- Figures
- Tables
- 1. About This Manual
- Copyright © 2012 by Despatch Industries.
- Printed and bound in the United States of America.
- Table 1. Operating/Environmental Conditions (For indoor use).
- 2. Safety
- 2.1. Safety Information
- 2.1.1. Lockout
- 2.1.1.1. Lockout Requirements
- 2.1.1.2. Lockout Procedure
- 2.2. Provisions for Lifting and Carrying
- 2.3. Maintenance
- 2.4. Electrical Power
- 2.5. Fire
- 2.6. Equipment Lockout Requirements
- 2.6.1. Emergency Stop
- 2.7. Disconnecting Devices
- 2.7.1. Power Requirements
- 2.7.2. Disconnecting Hard-Wired Units
- 2.7.3. Disconnecting Corded Units
- 2.7.4. Disconnecting Units with Optional Disconnect Switch
- 3. Theory of Operation
- Table 2. Control Instrument Explanations.
- Review the Protocol 3 Controller Owner’s Manual for more information.
- Table 3. High Limit Instrument Explanations.
- 4. Assembly & Setup
- 5. Operation
- Table 4. Control Instrument Operating Parameters.
- Table 5. Control Instrument Set-up Parameters.
- Refer to Table 5 to change tuning parameters, if necessary.
- Table 6. High Limit instrument Setup Parameters.
- 6. Maintenance
- 6.1. Checklist
- 6.2. Lubrication
- 6.3. Cleaning and Decontamination
- 6.3.1. Cleaning the LBB Series Oven
- 6.3.2. Decontaminating the LBB Series Oven
- 6.4. Routine Tests
- 6.4.1. Test Control Instrument
- 6.4.2. Test High Limit Instrument
- 6.5. Replacement Parts
- 6.5.1. Replace the Control Instrument
- 6.5.2. Replace High Limit Instrument
- 6.5.3. Replace (Optional) Protocol 3 Controller
- 6.5.4. Replace Heater Unit
- 6.5.5. Replace Fan Motor
- 7. Troubleshooting
- Table 7. Common Technical Issues and Remedies.
- 8. Appendices
- Each oven schematic (Section 0) shows the wiring for this option.
- Each oven schematic (Section 8.4) shows the wiring for this option.
- Each oven schematic (Section 0) shows the wiring for this option.
- Table 8. LBB1-23A-1 Parts.
- Table 9. LBB1-23B-1 Parts.
- Table 10. LBB1-43A-1 Parts.
- Table 11. LBB1-43B-1 Parts.
- Table 12. LBB1-69A-1 Parts.
- Table 13. LBB1-69B-1 Parts.
- Table 14. LBB2-12-1 Parts.
- Table 15. LBB2-18-1 Parts.
- Table 16. LBB2-27-1 Parts.

LBB Series Oven Owner’s Manual OPERATION
Version 15 31
Copyright © 2012 by Despatch Industries.
All rights reserved. No part of the contents of this manual may be reproduced, copied or transmitted in any form or by any
means including graphic, electronic, or mechanical methods or photocopying, recording, or information storage and
retrieval systems without the written permission of Despatch Industries, unless for purchaser's personal use.
For sample heating, close vent when no ventilation is required
5.3. Operating Procedure
Warning!
Do not use flammable solvent or other flammable material in this
oven. Do not process closed containers of any substance or
liquid in this oven because they may explode under heat.
For fastest oven heat-up time, close the exhaust vent. After the
desired temperature is reached, adjust the vent as needed.
5.3.1. Start Oven
1. Turn POWER ON (Figure 11). The circulating fan will start.
2. Set exhaust vent to desired opening by adjusting the damper at the rear of the oven (Figure
12).
a. Adjust the vent for maximum performance at various operating temperatures.
3. Enter desired setpoint on the High Limit instrument.
a. Set the High Limit instrument to a temperature 10° to 14°C higher than the desired
setpoint or to a temperature that should not be exceeded in the process.
b. If the LOCK on the High Limit instrument is lit, press and hold the RESET for four (4)
seconds to enable ▲ and ▼.
c. Press . HSP1 will light.
d. Use ▲ and ▼ to set the High Limit temperature.
e. Press RESET (or ) once to enter the value and return to the process variable PV
mode.
4. Enter setpoint on the Control instrument.
a. Press until the SP lights.
b. Use ▲ and ▼ to set the operating temperature.
c. Press to enter the setpoint.
d. Press again to display the process temperature.
5. Turn HEATER ON.
If the High Limit instrument is exceeded the heater shuts
down. Reset the High Limit instrument by pushing RESET on
the High Limit instrument.