Electromagnec Immunity Declaraon
The NIRA Skincare Laser is intended for use in the electromagnec environment specied below. The customer or the user of the NIRA Skincare Laser should assure that it is used in
such an environment.
Immunity test IEC 60601test level Compliance level Electromagnec environment guidance
Electrostac discharge (ESD)
IEC 61000-4-2
+8 kV contact
+15 kV air
Complies
Electrical fast transient/ burst
IEC 61000-4-4
+2 kV for power supply lines
+1 kV for input/output lines.
Complies
Mains power quality should be that of a typical home, commercial
or hospital environment.
Surge
IEC 61000-4-5
+1 kV dierenal mode
+2 kV common mode
Complies
Mains power quality should be that of a typical home commercial
or hospital environment.
Voltage dips, short interrupts and
voltage variaons on power supply
input lines
IEC 61000-4-11
<5 % UT (>95 % dip in UT)
For 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip in UT) for 5 sec
Complies
Power frequency (50/60 Hz)
magnec eld
IEC 61000-4-8
3 A/m Complies
Power frequency magnec elds should be at levels
characterisc of a typical locaon in a typical home, commercial or
hospital environment.
NOTE UT is the a.c. mains voltage prior to applicaon of the test level.