Use Instructions
21
Electromagnec Immunity Declaraon
The NIRA Skincare Laser is intended for use in the electromagnec environment specied below. The customer or the user of the NIRA Skincare Laser should assure that it is used in
such an environment.
Immunity test IEC 60601test level Compliance level Electromagnec environment – guidance
Electrostac discharge (ESD)
IEC 61000-4-2
+8 kV contact
+15 kV air
Complies
Electrical fast transient/ burst
IEC 61000-4-4
+2 kV for power supply lines
+1 kV for input/output lines.
Complies
Mains power quality should be that of a typical home, commercial
or hospital environment.
Surge
IEC 61000-4-5
+1 kV dierenal mode
+2 kV common mode
Complies
Mains power quality should be that of a typical home commercial
or hospital environment.
Voltage dips, short interrupts and
voltage variaons on power supply
input lines
IEC 61000-4-11
<5 % UT (>95 % dip in UT)
For 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT (>95 % dip in UT) for 5 sec
Complies
Power frequency (50/60 Hz)
magnec eld
IEC 61000-4-8
3 A/m Complies
Power frequency magnec elds should be at levels
characterisc of a typical locaon in a typical home, commercial or
hospital environment.
NOTE UT is the a.c. mains voltage prior to applicaon of the test level.