Use Instructions
16
SERVICE AND MAINTENANCE
The device does not require assembly or regular maintenance. Never aempt to service it in any manner. If you have any ques ons regarding
servicing or maintenance, please call Dermal Photonics Corporaon. The NIRA Skincare Laser will operate for more than one ye ar of recommended use.
CLEANING AND DISINFECTION
The interior of the p should be cleaned once a week with a coon swab. Gently place the coon swab into the opening on th e Laser Aperture to
clean the glass lens and remove any debris. If debris has fallen into the laser output window and it will not come out using the coon swab, gently
shake the device to let the debris fall out. Disinfect all surfaces with 70% alcohol soluon between each use.
ENVIRONMENT CONDITIONS
The NIRA Skincare Laser should be used in a comfortable environment.
Temperature: 15°C to 35°C (59°F to 95°F)
Relave Humidity: 30-75%
Atmospheric Pressure: 70-106 kPa
REQUIRED SHIPPING AND STORAGE CONDITIONS
The NIRA Skincare Laser should be stored in a dry place at normal room temperature.
Shipping Temperature: –40°C to 50°C (-40°F to 122°F)
Storage Temperature: –20°C to 35°C (-4°F to 95°F)
Relave Humidity: 10-100%