Electromagnec Immunity Declaraon
The NIRA BEAUTY is intended for use in the electromagnec environment specied below. The customer or the user should assu re that it is used in such an environment.
Immunity
test
IEC 60601 test level Compliance level Electromagnec environment – guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
10 V/m
80 MHz to 2,5 GHz
3 V
10 V/m
Portable and mobile RF communicaons equipment should be used no closer to any part of the NIRA BEAUTY including cables, tha n the
recommended separaon distance calculated from the equaon applicable to the frequency of the transmier. Recommended sepa raon
distance
d = 1.17
d = 1.17 80 MHz to 800 MHz
d = 2.33 800 MHz to 2,5 GHz
where P is the maximum output power rang of the transmier in was (W) according to the transmier manufacturer and d is the
recommended separaon distance in meters (m).
Field strengths from xed RF transmiers, as determined by an electromagnec site survey,
a
should be less than the compliance level in
each frequency range.
b
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situaons. Electromagnec propagaon is aected by absorpon and reecon from structures, objects and people.
a
Field strengths from xed transmiers, such as base staons for radio (cellular/cordless) telephones and land mobile radio , amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theorecally with accuracy. To assess the electromagnec environment due to xed RF transmiers, an electromagnec site survey should be considered. If the measured eld strength in the locaon in
which the NIRA BEAUTY is used exceeds the applicable RF compliance level above, the NIRA BEAUTY should be observed to verify normal operaon. If abnormal performance is observed, addional measures
may be necessary, such as reorienng or relocang the NIRA BEAUTY.
b