Use Instructions

21
Electromagnec Immunity Declaraon
The NIRA BEAUTY is intended for use in the electromagnec environment specied below. The customer or the user of the NIRA BEAUTY should assure that it is used in such an
environment.
Immunity test IEC 60601test level Compliance level Electromagnec environment guidance
Electrostac discharge (ESD)
IEC 61000-4-2
+8 kV contact
+8 kV air
Complies Floors should be wood, concrete or ceramic le. If oors are
covered with synthec material, the relave humidity should be at
least 30%
Electrical fast transient/ burst
IEC 61000-4-4
+2 kV for power supply lines
+1 kV for input/output lines.
Complies Mains power quality should be that of a typical home, commercial
or hospital environment.
Surge
IEC 61000-4-5
+1 kV dierenal mode
+2 kV common mode
Complies Mains power quality should be that of a typical home commercial
or hospital environment.
Voltage dips, short interrupts
and voltage variaons on
power supply input lines
IEC 61000-4-11
<5 % UT (>95 % dip in UT)
For 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25
cycles
<5 % UT (>95 % dip in UT) for 5 sec
Complies Mains power quality should be that of a typical commercial or
hospital environment. If the user of the NIRA BEAUTY requires
connued operaon during power mains
Interrupons, it is recommended that the NIRA BEAUTY be
powered from an uninterrupted power supply or a baery.
Power frequency (50/60 Hz)
magnec eld
IEC 61000-4-8
3 A/m Complies Power frequency magnec elds should be at levels
characterisc of a typical locaon in a typical home, commercial or
hospital environment.
NOTE UT is the a.c. mains voltage prior to applicaon of the test level.