User manual
Technical Information
246
mmol/L: 11.5 to 30 mmol/L in increments
of 0.5 mmol/L
Default: 14, 16.5, 22 and 28 mmol/L
Percent: 1% to 100% in increments of 1%
Default: 1%
Display Insulin on Board home screen: Yes or No
Default: No
Basal Pattern
Maximum basal rate: 0 (zero) or 0.05 u/hr to 35
u/hr in 0.05 u/hr increments
Default: 2 u/hr
Review or Edit Basal Pattern: Yes or No
Default: No
If Yes, user chooses number of patterns: 1 to 4
patterns may be displayed or not displayed.
Default: 1 pattern displayed
Pattern names: Pattern “X” (1, 2, 3, 4), Exercise
day, Weekday, Travel day, Weekend,
Monthly
Default: Pattern “X” (1, 2, 3, 4)
Temporary Rate
Temporary rate: Display in main menu: Yes or
No
Default: No
Temporary rate method: Percent or Units/Hr
Default: Percent
Active alarm (Give alert while temporary rate is
active): Yes or No
Default: Yes
If Yes, user chooses how oen to alert: 00:15 to
02:00 hours in 00:15 increments
Default: 00:15 minutes
End alarm (Give alert when complete): Yes or No
Default: Yes
Display Temporary Rate home screen: Yes or No
Default: No
Custom Temporary Rates:
Review or edit; active temporary rates
indicated by (up to 8 available,
programmed using CoZmanager® 2.0
soware).
Edit (Display in menu): Yes or No
Correction Factor: User may program 1 to 24
separate correction factors in increments of
30 minutes.
Default: A single correction factor,
programmable in the following increments:
mg/dl: 5 to 400 mg/dl in the following
increments:
•5to30mg/dlinincrementsof1mg/dl
•30to100mg/dlinincrementsof5mg/dl
•100to400mg/dlinincrementsof
10 mg/dl
Default: 40 mg/dl
mmol/L: 0.2 to 23 mmol/L in the following
increments:
•0.2to1.5mmol/Linincrementsof
0.05 mmol/L
•1.5to5mmol/Linincrementsof
0.25 mmol/L
•5.0to23mmol/Linincrementsof
0.5 mmol/L
Default: 2.25 mmol/L
Duration of insulin action: 02:00 to 08:00 hours
in 00:15 increments
Default: 03:00
Target blood glucose: User may program 1 to 24
separate targets in increments of 30 minutes
mg/dl: 80 to 200 mg/dl in 5 mg/dl
increments
Default: 150 mg/dl
mmol/L: 4.4 to 11.1 mmol/L in 0.1 mmol/L
increments
Default: 8.3 mmol/L
Edit Target BG with bolus: If enabled, allows
user to edit target BG while programming a
correction bolus: Yes or No
Default: No
Increase correction bolus based on BG: If
enabled, user can select up to four
thresholds which, if a reading which
exceeds those thresholds is entered, cause
the bolus to be increased by a specied
percent: Yes or No
Default: No
If Yes, resholds:
mg/dl: 210 to 540 mg/dl in increments of
10 mg/dl
Default: 250, 300, 400 and 500