Specifications
DELTA LTL2000 Retrometer 13
Test Printout
The Test Printout showing the measured Rl, the date and time for the test, the related status code and
a T# number showing the number of measurements in the Test log. The test printout can only follow after
a measurement performed with the <2'nd> + <TEST> key.
Signal levels and battery conditions are measured during the measurement and printed.
If any special condition exists (status > 0) then the status code is decoded and the conditions are printed
(see below), in the above example no special conditions exist. The measured signal and stray light levels
are shown in % of their full scale values. The battery voltage before and during measurement is shown.
Status code interpretation:
bit pattern value Description
Status Code: 00000001 1 Not defined
Status Code: 00000010 2 Stray Light Warning
Status Code: 00000100 4 Rl Log full Warning
Status Code: 00001000 8 Test Log full Warning
Status Code: 00010000 16 Low Battery Warning
Status Code: 00100000 32 High Zero signal Warning
Status Code: 01000000 64 High Signal with Lamp on Error
Status Code: 10000000 128 High Signal with Lamp off Error
eg. Status code 18 is composed of Low Battery Warning (16) + Stray Light Warning (2)