Instructions
HD32.8 - 35 - V1.3
Command Response Description
LDnn Print data stored in section nn. (nn= hexadecimal number 00-
63). If the section is empty: “-->No Log Data<--”
LEX
&
Cancel data stored in all sections.
LFnn Start on; 2007/01/01 00:37:32;
rec;000003
Print memory nn section status. The number, and storage start
date and time are displayed. (nn= hexadecimal number 00-63).
If the section is empty: “-->No Log Data<--”
LN 00 ; 01 ; 02 ; 03 ; . ; . ; . ; Print instrument memory map: If a section is allocated a
number is displayed, if it is free a point (.) is displayed.
P0 & Ping (locks the instrument keyboard for 70 seconds)
P1 & Unlocks the instrument keyboar
d
RA
Sample log = 30sec
Reading of LOG interval set
RL
Sample print = 0sec
Reading of PRINT interval set
S0 Printing of the current measurement of all inputs
S1 Printing of the REL current measurement of all inputs
WA#
&
Setting LOG interval.
# is a hexadecimal number 1…D that represents the position
of the interval in the list 2, 5, 10, …, 3600 seconds.
WC0
&
Setting SELF of
f
WC1
&
Setting SELF on
WL#
&
Setting PRINT interval.
# is a hexadecimal number 0…D that represents the position
of the interval in the list 0, 2, 5, 10, …, 3600 seconds.