Instructions

HD32.8 - 18 - V1.3
10. Press ENTER to confirm or ESC to exit without enabling the automatic start: In both cases,
you will return to the LOGGING menu.
11. Press SETUP to exit immediately from the main menu.
When the instrument starts automatically a LOG session, a beep is issued on each capture and the
blinking LOG message is shown at the top of the display.
Press MEM to stop the session before the set time.
To cancel the automatic start setting, use the Cancel auto start function as illustrated in the
following paragraph.
NOTE 1: The automatic logging session is started even when the instrument is off. If it is off
when the automatic logging session is started, the instrument is turned on few seconds before the
start time and, if powered by the mains, it remains on at the end of logging. If it is powered by the
battery, it is turned off when idle for some minutes at the end of the logging session.
NOTE 2: Use of the MEM key after setting the automatic start and stop.
If the automatic start and stop have been set and, before the automatic start, you press the MEM
key, the instrument starts recording at once. The next automatic start command is ignored, the
recording ends at the automatic stop command.
If the automatic start and stop have been set and, after the automatic start, you press the MEM
key, the instrument stops recording. The next preset automatic stop command is cleared.
Automatic Start
Automatic Stop MEM key