HD32.8 - 15 - V1.3
4.2 LOGGING MENU
To access the Logging menu, open the main menu using the SETUP key. Using the ▲▼ ◄►
arrows, select Logging and confirm with ENTER.
The parameter setting submenu for the logging sessions will be displayed.
You can set the data capture frequency (Log interval) and the automatic l
gg g
oin start
(
/sop m
Startt tie).
Th pt
e ca
r nt r v l t h m f or l l pr ob
u e i e a is e sa e a es.
With the “Log File Manager” you can handle the stored data sessions: printing, erasure.
4.2.1 Log Interval
Use this item to set the LOG interval (interval between two subsequent sample records): To enter
this setting, proceed as follows.
Once you have accessed the LOGGING submenu (previous par.) use the arrows to select Log
Interval:
1. Use the arrows to select the interval duration from a minimum of 2 seconds to a
maximum of one hour;
2. Press ENTER to confirm and return to the Logging menu;
3. Press ESC to return to the Logging menu without making any change;
4. Press ESC again to return to the main menu;
5. Press SETUP to exit immediately from the menu.
These are the available values:
2, 5, 10, 15, 30 seconds, 1, 2, 5, 10, 15, 20, 30 minutes, and 1 hour.
4.2.2 Self Shut-off mode
The Self shut-off mode item controls the instrument’s automatic turning off during logging,
between the capture of a sample and the next one. When the interval is lower than 60 seconds, the
instrument will always remain on. With intervals greater than or equal to 60 seconds, it is possible
to turn off the instrument between loggings: it will turn on 15 seconds before the sampling and will
turn off immediately afterwards, thus increasing the battery life.