Instructions
HD32.3  - 3  -  V1.4 
1. GENERAL FEATURES 
HD32.3 is designed for: 
• Analysis of hot environments through WBGT index (Wet Bulb Globe Temperature) in pres-
ence or absence of solar radiation. 
• Analysis of moderate environments through PMV index (Predicted Mean Vote) and PPD in-
dex (Predicted Percentage of Dissatisfied). 
The instrument is provided with three inputs for probes with SICRAM module: the probes have an 
electric circuit that communicates with the instrument; the sensor calibration data are saved in its 
permanent memory. 
All the SICRAM probes can be inserted in one of the inputs: they are automatically recognized 
when you switch the instrument on. 
The main features of the instrument are: 
•  Logging: acquisition and memorization of data inside the instrument. Memory capacity: 64 
different logging sections, with the possibility to set the acquisition interval of the samples. 
You can set the duration of the memorization and, with auto-start function; it’s possible 
to set the starting and finishing date and the hour of data memorization. 
•  The measurement unit of the temperature visualized sizes: °C, °F, °K. 
•  The system date and hour. 
•  The visualization of the maximum, minimum and medium statistic parameters and their 
cancellation. 
•  The transfer speed of the data through the serial port RS232. 
NOTE: The set acquisition interval is the same for all the probes connected to the instrument. 
HD32.3 instrument can contemporarily detect the following sizes: 
•  Globe thermometer temperature Tg with TP3276.2 or TP3275 probe. 
•  Wet bulb temperature with natural ventilation Tn with HP3201.2, HP3201 or TP3204S 
probe. 
•  Ambient temperature T with TP3207.2 or TP3207 probe. 
•  Relative Humidity RH and ambient temperature T with HP3217.2 or HP3217R probe. 
•  Air speed Va with AP3203.2 or AP3203 probe. 
On the basis of the detected measurements, HD32.3 with TP3207.2 (or TP3207), HP3276.2 (or 
HP3275) and HP3201.2 (or HP3201 / TP3204S) can calculate: 
•  WBGT(in) index (Wet Bulb Glob Temperature: wet bulb and globe thermometer tempera-
ture) in absence of solar irradiation. 
•  WBGT(out) index (Wet Bulb Glob Temperature: wet bulb and globe thermometer tempera-
ture) in presence of solar irradiation. 
On the basis of the detected measurements, HD32.3 with HP3217.2 (or HP3217R), HP3276.2 (or 
HP3275) and AP3203.2 (or AP3203) probes can calculate: 
•  Medium radiant temperature Tr. 
•  PMV index (Predicted Mean Vote). 
•  PPD index (Predicted Percentage of Dissatisfied). 










