Instructions

HD2402 - 42 - V1.3
o L
R
(11mrad) [W/(sr)] radiance calculated assuming that the source angular
dimension is 11mrad.
o L
R
(real) [W/(m² sr)] real radiance calculated starting from the entered geomet-
ric parameters.
o t_lim [s] exposure time limit, that is the time period during which a source can
be observed without damages to the visual system. This value is calculated with
the following procedure:
if Alpha > 100mrad it is assumed that C
α
= 100
if L
R_real
6
10
4
t > 10s (j limit satisfied)
if 6
10
4
L
R_real
8.89
10
6
t_lim = (5
10
5
/L
R_real
)
4
[s] (k limit satisfied)
if 8.89
10
6
L
R_real
t_lim < 10μs (l limit)
if 11mrad < Alpha < 100mrad it is assumed that C
α
= Alpha
if L
R_real
6
10
6
/Alpha t > 10s (j limit satisfied)
if 6
10
6
/Alpha L
R_real
8.89
10
8
/Alpha t_lim = (5
10
7
/(L
R_real
Alpha))
4
[s]
(k limit satisfied)
if 8.89
10
8
/Alpha L
R_real
t_lim < 10μs (l limit)
if Alpha < 11mrad it is assumed that C
α
= 11
if L
R_11
5.45
10
5
t > 10s (j limit satisfied)
if 5.45
10
5
L
R_11
8.08
10
7
t_lim = (4.54
10
6
/(L
R_11
))
4
[s] (k limit satis-
fied)
if 8.08
10
7
L
R_real
t_lim < 10μs (l limit)
The exposure limit time in seconds is displayed aside radiance values.
If the Include in Report box is selected, table data will be included in the final report.