Instructions
HD2402 - 27 - V1.3
6.4.3 Table a. 180-400
Table a. 180-400 describes the exposure limit for parameter “a”.
The displayed data are:
• Effective irradiance value E
eff
[W/m
2
] (corresponds to the instant value if a sin-
gle sample is considered, or to the mean value if several samples are consid-
ered).
• Effective dose value H
eff
[J/m
2
] regarding the considered samples.
• Time in hours/minutes/seconds for exceeding 30 J
eff
/m
2
maximum daily dose
calculated starting from the irradiance mean value.
If the Include in Report box is selected, the data in the table will be included in the fi-
nal report.
6.4.4 Table b. 315-400
Table b. 315-400 shows the exposure limit for parameter “b”.
The displayed data are:
• Irradiance value E
UVA
[W/m
2
] (corresponds to the instant value if a single sam-
ple is considered, or to the mean value if several samples are considered).
• Dose value H
UVA
[J/m
2
] regarding the considered samples.
• Time in hours/minutes/seconds for exceeding 10000 J
UVA
/m
2
maximum daily
dose calculated starting from the irradiance mean value.
If the Include in Report box is selected, the data in the table will be included in the fi-
nal report.