Data Sheet

CP24H
SWD05
VTRAP20
HD 2402
The HD2402 is a portable photo-radiometer data logger for the measurement of non-
coherent optical radiation in compliance with the European Directive 2006/25/EC and
the legislative decree n. 81 of April 9
th
2008.
The instrument is equipped with a series of sensors to cover different spectral portions and a
small laser suitable to indicate the analyzed source.
The various sensors work in the following spectral ranges:
Photometric sensor for measuring illuminance (lux meter) in the spectral range 380...780 nm.
Radiometric sensor for the UV band (220...400 nm) with spectral weighting factor S(λ).
Radiometric sensor for UVA band (315...400 nm).
Radiometric sensor for the band 400...700 nm (blue) with spectral weighting factor B (λ).
Radiometric sensor for the IR band (700...1300 nm) with spectral weighting factor R(λ).
Thermopile sensor for the measurement of irradiance in the infrared spectral range
400...2800 nm.
The HD2402 can be power supplied either by the connection to a PC, receiving power supply
directly from the USB port of the PC, or by an external power supply with USB output (code
SWD05). The connection cable CP24H is equipped with an M12 connector on the instrument
side and a USB type connector for the PC side or to the power supply SWD05 side.
By using the software DeltaLog13 from the version 1.0.1.0 and a PC, the HD2402 can
be confi gured (calendar, date, time, starting time and duration of the logging) as well as
performing the download and the analysis of the data stored and the acquisition of data in real
time. Once confi gured, the data logger can be disconnected from the PC and connected to its
external power supply for the acquisition and storage of data according to the programmed
settings.
Instrument specifi cations
Instrument
Dimensions
(Length x Width x Height) 69x69x155 mm
74x74x155 mm with protective shell
Weight 500 g
Materials Aluminium alloy
Protective shell Rubber
Operating conditions
Working temperature -5 … 50°C
Storage temperature -25 … 65°C
Working relative humidity 0 … 85% RH no condensation
Protection degree IP 64
Power supply
Power adapter (cod. SWD05) 5Vdc/1A
Stored data security unlimited
Serial interface: output for connection to the PC by using
the USB cable CP24H
Storage capacity: 96,000 recordings, corresponding to
approximately 26 hours of continuous
data acquisition.
Storage interval: xed at 1 second.
Measuring ranges
Measurement of the illuminance in the
spectral range 380...780 nm
1.0 ... 399.9 lux
0.010·10
3
... 3.999·10
3
lux
0.10·10
3
... 39.99·10
3
lux
1.0·10
3
... 399.9·10
3
lux
Measurement of the UV radiation in the
spectral range 220...400 nm with
spectral weighting factor S(λ)
0.10·10
-3
... 39.99·10
-3
W/m
2
1.0·10
-3
... 399.9·10
-3
W/m
2
0.010 ... 3.999 W/m
2
0.10 ... 39.99 W/m
2
Measurement of the ultraviolet radiation
in the spectral UVA range (315...400 nm)
0.010 ... 3.999 W/m
2
0.10 ... 39.99 W/m
2
1.0 ... 399.9 W/m
2
0.010·10
3
... 3.999·10
3
W/m
2
Measurement of the radiation in the spectral
range 400...700 nm (blue) with
spectral weighting factor B(λ)
1.0·10
-3
... 399.9·10
-3
W/m
2
0.010 ... 3.999 W/m
2
0.10 ... 39.99 W/m
2
1.0 ... 399.9 W/m
2
Measurement of infrared radiation in the
spectral fi eld 700...1300 nm, with
spectral weighting factor R(λ)
0.010 ... 3.999 W/m
2
0.10 ... 39.99 W/m
2
1.0 ... 399.9 W/m
2
0.010·10
3
... 3.999·10
3
W/m
2
Measurement of infrared radiation, spectral
range 400...2800 nm
0.010·10
3
... 3.999·10
3
W/m
2
HD2402
INCOHERENT OPTICAL RADIATION MONITORING
ORDERING CODES
HD2402: Multi-sensor instrument, data logger, for measuring noncoherent optical radiation.
Equipped with: DeltaLog13 software (version 1.0.1.0) downloadable from Delta OHM
website, to download, monitor and process the data on a personal computer, hardware
key CH20-ROA to enable the software, CP24H connection cable, SWD05 external
power supply, VTRAP20 tripod, manual, carrying case.
Accessories:
CH20-ROA: Hardware key for PC with Windows
®
operating systems. Inserted into a USB port
enables the use of PC software DeltaLog13 with the instrument HD2402.
CP24H: Connection cable to a PC or to the external power supply. M12 connector on the
instrument side and USB type A- connector on the PC / Power Supply side.
SWD05: Stabilized external power supply 100...240Vac/5Vdc-1A. Output with USB connector
type A.
VTRAP20: Tripod to fi x the instrument, maximum height 270 mm.

Summary of content (1 pages)