Instructions
- - 
18 
9.2  TECHNICAL  CHARACTERISTICS  OF  PHOTOMETRIC  AND  RADIOMETRIC  PROBES  COMPLETE 
WITH SICRAM MODULE EQUIPPED WITH THE INSTRUMENT  
ILLUMINANCE measurement probe LP 471 PHOT complete with SICRAM module and equipped with the in-
strument 
Measurement range (lux): 
0.01…199.99 
…1999.9 
…19999 
…199.99 10
3
Resolution (lux): 
0.01 
0.1 
1 
0.01 10
3
Spectral range: 
in agreement with standard photopic curve V( ) 
α (temperature coefficient) f
6
(T): 
<0.05% K 
Calibration uncertainty: 
<4% 
f'1 (in agreement with photopic response V( )): 
<6% 
f
2 
(response according to the cosine law): 
<3% 
f
3
 (linearity): 
<1% 
f
4
 (instrument reading error): 
<0.5% 
f
5
 (fatigue): 
<0.5% 
Class: 
B 
Drift after 1 year: 
<1% 
Functioning temperature: 
0…50  C 
Reference Standard 
CIE No. 69 – UNI 11142 
LUMINANCE measurement probe LP 471 LUM 2 complete with SICRAM module and equipped with the in-
strument 
Measurement range (cd/m
2
): 
0.1…1999.9 
…19999 
…199.99 10
3
…1999.9 10
3
Resolution (cd/m
2
): 
0.1 
1 
0.01 10
3
0.1 10
3
Optical angle: 
2  
Spectral range: 
in agreement with standard photopic curve V( ) 
α (temperature coefficient) f
6
(T): 
<0.05% K 
Calibration uncertainty: 
<5% 
f'1 (in agreement with photopic response V( )): 
<8% 
f
3
 (linearity): 
<1% 
f
4
 (instrument reading error): 
<0.5% 
f
5
 (fatigue): 
<0.5% 
Class: 
C 
Drift after 1 year: 
<1% 
Functioning temperature: 
0…50  C 
Reference Standard 
CIE No. 69 – UNI 11142 
Typical response curve 
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
380 430 480 530 580 630 680 730 780 830
Lambda (nm)
Relative spectral response