The HD2302.0 is a portable instrument with a large LCD display. It measures illuminance,
luminance, PAR and irradiance (across VIS-NIR, UVA, UVB and UVC spectral regions or
measurement of irradiance effective according to the UV action curve). The probes are equipped
with the SICRAM automatic detection module: in addition to detection, the unit of measurement
selection is also automatic. The factory calibration data are already memorized inside the
instruments. The Max, Min and Avg function calculate the maximum, minimum or average
values. Other functions include: the relative measurement REL, the HOLD function, and the
automatic turning off that can also be excluded. The instruments have IP67 protection degree.
INSTRUMENT TECHNICAL CHARACTERISTICS
Instrument
Dimensions
(Length x Width x Height) 140x88x38mm
Weight 160g (complete with batteries)
Materials ABS
Display 2x4½ digits plus symbols - 52x42mm
Visible area: 52x42mm
Operating conditions
Operating temperature -5…50°C
Storage temperature -25…65°C
Working relative humidity 0…90%RH without condensation
Protection degree IP67
Power
Batteries 3 1.5V type AA batteries
Autonomy 200 hours with 1800mAh alkaline batteries
Power absorbed with the instrument off
20μA
Measuring unit lux - fcd - µmol/m
2
·s - cd/m
2
- W/m
2
- µW/cm
2
µW/lumen
Connections
Input module for the probes 8-pole male DIN45326 connector
Technical characteristics of photometric and radiometric probes equipped with
SICRAM module for the connection the instrument.
LP 471 PHOT probe for the measure of ILLUMINANCE
Measuring range (lux): 0.10…199.99 …1999.9 …19999 …199.99·10
3
Resolution (lux): 0.01 0.1 1 0.01·10
3
Spectral range:
in agreement with standard photopic curve V(λ)
α (temp. coeffi cient) f
6
(T)
<0.05%K
Calibration uncertainty: <4%
f’
1
(in agreement with photopic
response V(λ)):
<6%
f
2
(response according to the
cosine law):
<3%
f
3
(linearity): <1%
f
4
(instrument reading error): <0.5%
f
5
(fatigue): <0.5%
Class B
Drift after 1 year: <1%
Working temperature: 0…50°C
Reference Standards CIE n.69 - UNI 11142
Photometric probe for ILLUMINANCE measurement, spectral response in agreement with standard
photopic vision, diffuser for cosine correction. Measurement range: 0.10 lux…200·10
3
lux.
LP 471 LUM 2 probe for the measure of LUMINANCE
Measuring range (cd/m
2
): 1.0…1999.9 …19999 …199.99·10
3
…1999.9·10
3
Resolution (cd/m
2
): 0.1 1 0.01·10
3
0.1·10
3
Optical angle: 2°
Spectral range:
in agreement with standard photopic curve V(λ)
α (temp. coeffi cient) f
6
(T)
<0.05%K
Calibration uncertainty: <5%
f’
1
(in agreement with
photopic response V(λ)):
<8%
f
3
(linearity): <1%
f
4
(instrument reading error): <0.5%
f
5
(fatigue): <0.5%
Class C
Drift after 1 year: <1%
Working temperature: 0…50°C
Reference Standards CIE n.69 - UNI 11142
Photometric probe for LUMINANCE measurement, spectral response
in agreement with standard photopic vision, vision angle 2°.
Measurement range: 1.0 cd/m
2
…2000·10
3
cd/m
2
.
Typical response curve: LP 471 PHOT and LP 471 LUM2
HD2302.0
PHOTO-RADIOMETER