5. INSTRUMENT SIGNALS AND FAULTS 
The following table lists all of the indications displayed by the instrument in different operating and 
error situations: 
Display indications 
Explanation 
BATT TOO LOW 
CHNG NOW 
Indication of insufficient battery charge: it appears when you turn the 
instrument on. The instrument issues a long beep and turns off. 
Replace the batteries. 
CAL 
LOST 
Program error: it appears after turning on for a few seconds. 
Contact the instrument's supplier. 
PROB COMM LOST 
ERR 
This appears if the probe has already been detected by the instrument, 
but is disconnected. At the same time an intermittent beep is issued. 
FUNC CLRD 
Maximum  (MAX),  minimum  (MIN)  and  average  (AVG)  values 
cleared. 
NEW_PROB_DET 
New probe detected 
NO_PRBE_ 
SER_NUM 
The connected probe's serial number is absent 
OVER 
or 
- - - - 
Measurement overflow: indicates that the probe is measuring a value 
exceeding the measuring range. 
PLS_EXIT >>> FUNC 
RES_FOR_FACT 
ONLY 
Please exit using ESC >>> function reserved to factory calibration 
PRBE_SER #### #### 
Serial number #### #### of the connected probe 
PROB 
ERR 
A probe with SICRAM module has been inserted when not admissible 
for that instrument. 
PROB COMM LOST 
This appears if the probe has already been detected by the instrument, 
but is disconnected. At the same time an intermittent beep is issued. 
SYS 
ERR 
# 
Instrument management program error. 
Contact the instrument's supplier and communicate the numeric code # 
reported by the display. 