Instructions

- -
33
Confirm with ENTER. The following screen will appear:
PROBES CALIBRATION
CO CALIBRATION
CAL ZERO
Cal zero 0 ppm
<ESC> exit/cancel
<ENTER> confirm
Wait about 15 minutes before continuing.
Supply the gas adjusting the bottle flow meter to get a constant flow from 0.1 to 0.2 l/min.
Press ENTER and wait the two minutes necessary for calibration without changing the
working conditions.
At the end, close the bottle spigot and remove the cap from the CO sensor.
Insert the protection grid.
Sensitivity of the CO sensor:
The sensitivity in nA/ppm of the CO sensor is already set at the factory. If you need to chage it,
follow this procedure:
1. Use the arrows ▼▲ to select the 2) Calibration CO item:
PROBES CALIBRATION
1) Calibration CO
2
2) Calibration CO
3) Calibration RH
▼▲ select
<ESC> exit/cancel
<ENTER> confirm
2. Confirm with ENTER. The following screen will appear:
PROBES CALIBRATION
CALIBRATION CO
1) Cal zero
2) Set sensitivity
▼▲ select
<ESC> exit/cancel
<ENTER> confirm