Instructions
- -
31
4.5.2 CO Calibration
You can calibrate the zero of the CO sensor in clean air (the CO concentration is lower than 0.1
ppm outdoor) or using nitrogen bottles (code MINICAN.12A).
Use the arrows ▲▼ to select the 2) Calibration CO item:
PROBES CALIBRATION
1) Calibration CO
2
2) Calibration CO
3) Calibration RH
▼▲ select
<ESC> exit/cancel
<ENTER> confirm
Confirm with ENTER. The following screen will appear:
PROBES CALIBRATION
CALIBRATION CO
1) Cal zero
2) Sensitivity set
▼▲ select
<ESC> exit/cancel
<ENTER> confirm
Confirm with ENTER. The following screen will appear:
PROBES CALIBRATION
CALIBRATION CO
CAL ZERO
Cal zero 0 ppm
<ESC> exit/cancel
<ENTER> confirm
Place the instrument in a clean air environment (CO concentration is lower than 0.1 ppm outdoor).
Turn the instrument on and wait at least 15 minutes so as to stabilize the measurement. Now press
ENTER and wait the two minutes necessary for calibration without changing the working
conditions.