Instructions

- -
17
4.2 LOGGING MENU
Enter the main menu by pressing MENU;
Use the arrows ▲▼ to select Logging;
Press ENTER: The parameter setting submenu for the logging sessions (to be captured) will
be displayed.
LOGGING MENU
1) Log frequency
2) Auto switch off
3) Start/Stop Log
4) Start Log Erase
5) Log File Manager
▲▼ select
<ENTER> confirm
4.2.1 Log Interval
Use this item to set the LOG interval (interval between two subsequent sample captures): To enter
this setting, proceed as follows:
Once you have accessed the LOGGING submenu (previous par.) use the arrows ▲▼ to select Log
frequency:
LOGGING MENU
LOG FREQUENCY
Insert interval
of memorization
h:mm:ss (1h max)
0:00:15
▲▼ set
<ENTER> confirm
1. Use the arrows ▲▼ to select the interval duration from a minimum of 15 seconds to a maximum
of one hour;
2. Press ENTER to confirm and return to the Logging menu;
3. Press ESC to return to the Logging menu without making any change;
4. Press ESC again to return to the main menu;
5. Press MENU to exit immediately from the menu.
These are the available values: 15 seconds - 30 seconds - 1 minute - 2 minutes - 5 minutes - 15
minutes - 20 minutes - 30 minutes - 1 hour
Lo
gg
in
g
interval
Storage capacity
Lo
gg
in
g
interval
Storage capacity
15 seconds About 11 days and 17 hours 15 minutes About 1 year and 339 days
30 seconds About 23 days and 11 hours 20 minutes About 2 years and 208 days
1 minute About 46 days and 22 hours 30 minutes About 3 years and 313 days
2 minutes About 93 days and 21 hours 1 hour About 7 years and 261 days
5 minutes About 234 days and 17 hours