Instructions

HD208 - 6 - V1.4
3.1 BATTERY
The instrument uses a 3.6 V non-rechargeable lithium-thionyl chloride (Li-SOCl
2
)
battery AA size. To connect the battery, or to replace a dead battery with a new one,
proceed as follows:
1. Unscrew the 4 screws on the back of the case and remove the back cover.
2. In case of replacement, disconnect the battery connector from the circuit board and
replace the battery with a new one of the same type.
3. Connect the battery connector to the circuit board, observing the correct polarity.
The connector is equipped with a polarization key that prevents the possibility of a
wrong insertion of the connector.
4. Close the case by fixing the 4 rear screws (pay attention to the correct placement
of the battery, not to hinder the closing of the case).
The battery symbol at the bottom left of the display lights up when the battery is low;
in this case, replace the battery as soon as possible.
3.2 CONFIGURATION
The instrument parameters (date/time, logging parameters, alarm thresholds, quanti-
ties to be acquired) are configurable by connecting the instrument to a PC and using
the HD35AP-S application software (please see the instructions of the software) or, al-
ternatively, a specially designed PDF form included in the HD35AP-S software pack-
age (file HD208 configuration.pdf in folfer PDF/Language/Config_HD208).
The use of the PDF form must be enabled with HD35AP-S software. For the use of the
PDF form, please see the HELP button in the form.
Batter
y
Batter
y
connector