Data Sheet

6
Example of a PDF report – Summary data
PDF REPORT
At the end of each logging session, the data logger automatically generates a PDF report, which can then be copied to the PC via the USB port of the instrument. When generating the
report, the display of the instrument shows PdF.
The report includes the graphs of the detected quantities and information about the logging session: logging start and stop time, logging interval, number of samples acquired, alarm
thresholds, minimum, average and maximum of each detected quantity.
The report includes the calculation of the Mean Kinetic Temperature (MKT). The Mean Kinetic Temperature is an evaluation index of the cold chain used in the pharmaceutical field,
and is calculated according to the Haynes equation as a function of all the temperature measurements acquired during the logging session. The Mean Kinetic Temperature is used
to evaluate temperature fluctuations experienced by a biological substance during storage or transport, and corresponds to the storage temperature that, if maintained constant,
produces on the biological substance the same effects of the actual temperature changes recorded in the time period considered (i.e. the duration of the logging). You can set the
value of the activation energy, parameter necessary for the calculation of MKT.