User manual

4
Thermocouple (TC) S type 9
0 to 1700
o
C
Thermocouple (TC) R type 8
0 to 1700
o
C
Thermocouple (TC) N type 7
-200 to 1300
o
C
Thermocouple (TC) E type 6
0 to 600
o
C
Thermocouple (TC) T type2 5
-20.0 to 400.0
o
C
Thermocouple (TC) T type1 4
-200 to 400
o
C
Thermocouple (TC) J type2 3
-20.0 to 400.0
o
C
Thermocouple (TC) J type1 2
-100 to 850
o
C
Thermocouple (TC) K type2 1
-20.0 to 500.0
o
C
Thermocouple (TC) K type1 0
-200 to 1300
o
C
Thermocouple (TC) L type 16
-200 to 850
o
C
Thermocouple (TC) U type 17
-200 to 500
o
C
Thermocouple (TC) Txk type 18
-200 to 800
o
C
Input Error Indication
Setting value Temperature sensor is not connected Measured temperature value exceeds the temperature range Unknown input
PV
SV
Operation
There are three modes of operation: operation, regulation and initial setting. When power is applied, controller gets into the operation mode. Press the key to
switch to regulation mode. If the
key is pressed for more than 3 seconds, controller will switch to the initial setting mode. Pressing the key while in the
regulation mode or initial setting mode, forces the controller to return to the operation mode.
PV/SV: Sets the temperature set point and displays the temperature process value. Use the
and keys to set the temperature set point.
Setting method: While in any function mode, press the
key to select the desired function and use the and keys to change settings. Press key to
save the changes.
The next flow chart shows how to switch for settings and internal functions:
Regulation
Mode
Operation
Mode
Initial Setting
Mode
Press key less than 3 sec
Press key
Press key
Press key more than 3 sec