Oxygen Equipment User Manual
23
DL-00096 Rev. 7
Immunity test 
Electrostatic 
discharge(ESD) 
IEC61000-4-2 
Electricfast 
transient/burst 
IEC61000-4-4 
Surge 
IEC61000-4-5
IEC 60601 test 
level 
±6kVcontact
±8kVair
±2kVfor
power supply 
lines 
±1kVfor
input/output
lines
±1kV
differential mode 
±2kVcommon 
mode
Electromagnetic 
environment/guidance 
Floorsshouldbewood,
concrete,orceramictile.
Iffloorsarecoveredwith
synthetic material, the 
relativehumidityshould
beatleast30%.
Mainspowerquality
should be that of a typical 
commercial or hospital 
environment. 
Mainspowerquality
should be that of a typical 
commercial or hospital 
environment.
Guidance and manufacturer’s declaration: electromagnetic immunity
ThePortableOxygenConcentratorisintendedforuseintheelectromagnetic 
environmentspecifiedbelow.Thecustomerortheuseroftheconcentratorshould
assurethatitisusedinsuchanenvironment.
Compliance 
level 
±15kVcontact
±8kVair
±2kVfor
power supply 
lines 
±1kVfor
input/output
lines
±1kV
differential mode 
±2kVcommon 
mode
DL-00096 rev 7.indd 23 6/1/2009 4:48:35 PM