Oxygen Equipment User Manual
19
DL-00096 Rev. 7
Message 
Batterylow 
Fullycharged 
Checkingpower 
Nobattery 
BattLifeNN% 
Charging:NN% 
Charging 
RTXXXXX 
BreathrateXX 
Alarm cleared
Description 
Estimatedbatterylifeless
than17minutes.Theunit
will also generate a high 
priority alarm as described 
above. 
Displayedasthebattery
menu item when the  
batteryis100%charged.
Displayedasthebattery
menu item when the  
battery status has recently 
changed and the unit  
software is analyzing the 
changeinstatus.
Displayedasthebattery
menu item when there are 
no communications with 
thebattery. 
Displayedpercentageof
battery charge if at least 
10%andthereisno 
externalpowerconnected.
NN%displaysthecur-
rentbatterychargelevel.
Displayedwhenbattery
chargeisgreaterthan10%
butlessthan100%and
thereisexternalpower
connected.
Batterychargeislessthan
10%andthereisexternal
powerconnected.
Non-resettabletotalelapsed
timeindicator(hours).
Thepatient’saveragebreath
rate whenthedeviceis
deliveringthemaximum
amountofoxygenand
the bolusisreduced.Ifno
breaths are detected, the 
most recent breath rate is 
shown.
Apreviouslysetalarmhas
beenautomaticallycleared.
Action 
Chargethebatterypack
by plugging in to power 
supply. 
Noactionisrequired. 
Noactionisrequired. 
Verifythatthebat-
tery pack is cor-
rectlyinstalled.Contact
CustomerServiceifthe
battery is fully inserted 
and the message 
continues to be displayed 
longerthan30seconds.
Messageisdisplayed
when mode key is 
pressed. 
Messageisdisplayed
when mode key is 
pressed. 
Messageisdisplayed
when mode key is 
pressed.
Noactionrequired. 
Reduceactivitylevel.Be
sureairinlet/outlethas
notbeenblocked.
Note: The message will 
automatically clear when 
the device returns to normal 
operation.
Noactionrequired.
Other Messages, continued
Note: The message will be 
automatically cleared when 
plugged in to power supply.
DL-00096 rev 7.indd 19 6/1/2009 4:48:34 PM