Oxygen Equipment User Manual
18
DL-00096 Rev. 7
Alarm Message  
ReplaceC-filtr 
ReplaceP-filtr 
WashI-filtr 
Replace batt 
XX:Service*
*Value:71-99
Description 
Compressorfilterneeds
replacement. 
Patientfilterneeds 
replacement. 
Inputfilterneedstobe
cleaned. 
Mainbatteryhasexceeded
300rechargecycles.
Servicerequired.
Action 
Replacefilter.Pressand
hold mute key for more 
than4secondstoreset
thehourcounter.
Replacefilter.Pressand
hold mute key for more 
than4secondstoreset
thehourcounter.
Washinputfilter.Press
and hold mute key for 
morethan4secondsto 
resetthehourcounter.
ContactCustomer
Serviceforreplacement.
Note: This alarm will be 
automatically cleared once 
a new battery is inserted. 
Battery will continue to 
function after 300 recharge 
cycles, but for a shorter time 
than a new battery.
ContactCustomer
Service.
Low Priority Alarms
Other Messages
Message 
Powerremoved 
SWVerN.NNN 
Self test 
Shutting off
Externalpower 
Description 
Externalpowerhasbeen
disconnected; unit is now 
runningonbattery.
Displayofsoftwareversion
during startup and calibra-
tion.Thisincludestestof
beeperandLEDs.
Indicatesunitisgoing
throughitstestatstartup.
Displayedforapproximately 
10seconds.
Displayedwhileunitgoes
throughitspower-down
sequence.
Externalpoweris 
connected.
Action 
Noactionisrequired. 
Noactionisrequired. 
Noactionisrequired. 
Noactionisrequired.
Noactionisrequired.