Oxygen Equipment User Manual
17
DL-00096 Rev. 7
Alarm Message  
CheckVents 
LowBattery 
XX:Service!* 
*Value:21-50
Description 
Deviceisunabletomaintain 
oxygenpurity.
Estimatedbatterylifeless
than17minutes. 
Servicerequired.
Action 
Besureairinlet/outlet
hasnotbeenblocked.If 
alarm persists, contact 
CustomerService.
Chargethebatterypack
by plugging in to power 
supply.
Note: The message will be 
automatically cleared when 
plugged in to power supply. 
ContactCustomer
Service.
High Priority Alarms
Note: These alarms will allow the Portable Oxygen Concentrator to continue operating.
Alarm Message  
CheckCannula 
XX:Service*
*Value:51-70
Description 
Nobreathdetectedfor5
minutes. 
Servicerequired.
Action 
Checkthecannulacon-
nection.Besureto
breathethroughnose.If
alarm persists, contact 
CustomerService.
Note: The message will be 
automatically cleared when 
breathing is detected. 
ContactCustomer
Service.
Medium Priority Alarms