Oxygen Equipment User Manual
16
DL-00096 Rev. 7
IfthePortableOxygenConcentratordetectsanalarmcondition,itwillindi-
catethealarmvisuallyandaudiblywithin10seconds.Therearefourlevels
ofalarms:criticalhighpriority,highpriority,mediumpriority,andlowpriority.
Eachisindicateddifferentlybythebacklitdisplay;green,yellow,andred
LEDs;andbuzzer,asindicatedbelow.Ineachcase,thealarmmessageand
powerstatuswilloverridethecurrentdisplay.
Note: All alarm conditions and parameters are factory preset; conditions and parameters cannot be 
changed or adjusted by the user.
Alarm Status  
Criticalhigh 
priority 
High priority 
Mediumpriority 
Lowpriority
Audible Tone  
Ten beeps per burst, burst 
repeatsevery3seconds 
Ten beeps per burst, burst 
repeatsevery3seconds
Three beeps per burst, 
burstrepeatsevery8 
seconds 
Three beeps per burst, 
burstrepeatsevery 
10minutes
Visual Indicator 
SolidredLED 
(anddeviceshuts 
offautomatically)
FlashingredLED 
FlashingyellowLED 
SolidyellowLED
Note: If two alarm conditions exist at the same time, the highest priority alarm is indicated. If two or 
more alarm conditions of equal priority exist at the same time, the most recent one will be displayed.
Note: The most recent alarms indicated by the device are logged for reference by service personnel. 
This log is maintained even if the device is powered down or if power is lost for any other reason.
Note: If the mute key is pressed prior to an alarm condition (for example, to mute the device in a 
movie theater), critical high priority and high priority alarms will override the mute function; medium 
and low priority alarms will be muted for eight hours from the time the key was pressed. Press the 
mute key off to display the last highest priority alarm. Press the mute key on again to reset the eight-
hour timer.
ALARMS
WhenthePortableOxygenConcentratorsoundsanalarm,acorresponding
messagewillbedisplayedonthescreen.Takeappropriateactionasdirected
inthechartsbelow.
Critical High Priority Alarms
Note: These alarms will disable the Portable Oxygen Concentrator immediately.
Alarm Message  
ChargeBattery 
XX:Service!* 
*Value:01-20
Description 
Batteryneedscharging. 
Servicerequired.
Action 
Recharge the battery 
pack by plugging in to 
powersupply.Ensureall
connections are made 
securely.
ContactCustomer
Service.
Alarm Indicators
Mute Time 
20minutes 
20minutes 
8hours 
24hours