Use the keys to toggle between residential and commercial. After you have made your
selection, highlight DONE and press SELECT.
Please refer to Appendix A for a list of location definitions for residential jobs.
TO SET THE ID NUMBER:
From the “Create Job” screen, the ID field will display the last selected ID number. Press
SELECT on “ID” to enter the ID screen
Select any number between 0 and 999 as the ID of your job, using the keys. Press the
SELECT key after choosing the ID#.
TO SET TEMPERATURE CORRECTION:
Navigator Pro defaults to a temperature of 70ºF. As wood temperature increases, its electrical
resistance decreases and indicated moisture content rises. Lower wood temperatures result in
lower indicated moisture content. A correction is necessary if the wood temperature is outside
the range of 50ºF (10ºC) to 90ºF (32ºC). Set the temperature accordingly and the meter will
make the correction.
1. To change the temperature, use the keys from the CREATE JOB sub menu to
highlight the temperature, T: 70ºF.
2. Press the SELECT key to enter the COMP TEMP sub menu.
3. Using the keys, scroll up or down to a higher or lower temperature value.
Temperature values will change in 5º F increments.
4. To change between C and F, press the SELECT key when the COMP TEMP is highlighted.
TO DEFINE ROOM LOCATION AND MATERIAL DESIGNATION:
1. Using the keys on the CREATE JOB sub menu, highlight LOC DEF and press the
SELECT key.
2. Use the keys to choose different rooms and the keys to switch between
YES and NO options. YES means measurements will be taken in that room. NO means
that you are not going to test in that room. Residential jobs show a list of 32 rooms as
listed below: