Users Manual Part 7
Immunity Test IEC 60601 Test Level
Compliance
Level
Electromagnetic Environment -
Guidance
Magnetic Fields
generated by:
• Metal Detectors
• EAS Systems and
Tag Deactivators
(No Standard
Applied)
N/A
300 A/m
0.1 kHz —
3.5 kHz
50 A/m
10 kHz - 60
kHz
30 A/m
50 kHz - 135
kHz
Use in professional healthcare
facility and home healthcare
environments, excluding near high
frequency surgical equipment or
near MR machines.
RFID readers should be kept at
least 0.3 m [1 ft] away from the
Remunity
™
Pump for Remodulin
®
(treprostinil) Injection.
Hand held or walk through metal
detectors may be used near the
Remunity
™
Pump for Remodulin
®
(treprostinil) Injection, but avoid
prolonged exposure to them.
Step through retail anti-theft
detectors at a normal pace, but
avoid standing in them.
Table 30: Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
Appendix A
234 DKPI-09153-001-Rev. 1.4