Immunity Test IEC 60601 Test Level
Compliance
Level
Electromagnetic Environment -
Guidance
Electrical fast
transient/burst IEC
61000-4-4
+/-2 kV 100 kHz repetition
frequency
+/-2 kV 100
kHz
repetition
frequency
Mains power quality should be that
of a typical professional healthcare
facility and home healthcare
environments.
Surge IEC 61000-4-
5
± 0,5 kV, ± 1 kV line(s) to line(s)
± 0,5 kV, ± 1 kV, ± 2 kV line(s) to
earth
± 0,5 kV, ± 1
kV line(s) to
line(s)
± 0,5 kV, ± 1
kV, ± 2 kV line
(s) to earth
Mains power quality should be that
of a typical professional healthcare
facility and home healthcare
environments.
Table 30: Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
DKPI-09153-001-Rev. 1.4 231