Immunity Test IEC 60601 Test Level
Compliance
Level
Electromagnetic Environment -
Guidance
217 Hz PM
carrier
20 V/m
2.4 GHz – 2.5
GHz
111.5 kHz
PM carrier
Conducted RF IEC
61000-4-6
3 Vrms
(6 Vrms in ISM Bands)
150 kHz to 80 MHz
1 kHz 80% AM carrier
6 Vrms
150 kHz to 80
MHz
1 kHz 80%
AM carrier
Use in professional healthcare
facility and home healthcare
environments, excluding near high
frequency surgical equipment or
near MR machines.
Table 30: Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
Appendix A
230 DKPI-09153-001-Rev. 1.4