Users Manual Part 7

Guidance and Manufacturer’s Declaration - Electromagnetic
Immunity
Immunity Test IEC 60601 Test Level
Compliance
Level
Electromagnetic Environment -
Guidance
Electrostatic
discharge (ESD) IEC
61000-4-2
+/-2 kV, +/-4 kV, +/-6 kV, +/-8
kV contact
+/-2 kV, +/-4 kV, +/-8 kV & +/-
15 kV air
+/-2 kV, +/-4
kV, +/-6 kV,
+/-8 kV
contact
+/-2 kV, +/-4
kV, +/-8 kV &
+/-15 kV air
Use in professional healthcare
facility and home healthcare
environments.
Table 30: Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
Appendix A
226 DKPI-09153-001-Rev. 1.4