Users Manual Part 7

Quality of Service Provisions
Interruption or corruption of communication between the Remote Interface and Pump can lead to
interruptions in status updates and to making changes to therapy parameters. Momentary losses of
communication will be indicated by the status bar at the top of the Remote Interface screen being gray.
Interruption of communication for more than 2 hours will result in the Remote Interface declaring a No
Communication Attention Alarm.
Common consumer electronic devices that transmit in the same frequency band used by the Remunity
Pump for Remodulin
®
(treprostinil) Injection may prevent the Pump and Remote Interface from
communicating. Microwave ovens, Bluetooth devices, Wi-Fi networks and 2.4 GHz cordless phones, when
transmitting or receiving can cause interruption of communication between the Pump and Remote Interface.
This interference, however, will not cause any incorrect data to be sent and will not cause any harm to the
Remunity
Pump for Remodulin
®
(treprostinil) Injection.
In each of these cases, communication problems can usually be resolved by turning off or moving away from
other RF transmitting devices.
Appendix A
224 DKPI-09153-001-Rev. 1.4