Users Manual Part 7

Appendix A
Pump and Cassette Specifications Specification Details
Operating Environmental Range
Temperatures of 41 ˚F (5 ˚C) to 104 ˚F (40 ˚C)
Non-condensing humidity up to 90%
Pressures of 700 hPa - 1060 hPa
Infusion Accuracy 6%
Maximum Infusion Pressure < 113 kPa
Programmable Range
16 µL/h (0.016 mL) to 225 µL/h (0.225 mL), in 1µL/h (0.001
mL) increments
Pump Runtime with fully charged battery
Rate: 42 µL/h Runtime: >63.1 hours (Cassette volume
limited)
Rate: 225 µL/h Runtime: >11.8 hours (Cassette volume
limited)
Shelf Life
Refer to the Cassette packaging to determine Cassette
shelf life.
Expected Service Life
Cassette: 3 days
Pump: 3 years
Table 23: Pump andCassette Specifications
DKPI-09153-001-Rev. 1.4 219