Users Manual Part 7

Appendix A
Trumpet Curve
This trumpet curve shows the accuracy of the flow rate during a 4 hour and 46 minute period (100 deliveries) as
a function of an averaging window. The reported percent error deviation is calculated from the overall weight
increase over the full T2 time period. The data was collected at room temperature while delivering Remodulin
at a concentration of 1 mg/mL. The average shot cycle, for calculations per equations 21, 23 & 24 in sub-
clause 201.12.1.104 of IEC 60601-2-24: 2012, is taken as the average interval over the 100 shots in the
analysis period, T2.
DKPI-09153-001-Rev. 1.4 217