The pump can be worn in several ways: It can be placed in a pocket or attached to a belt by the included belt
clip accessory. There is no restriction on orientation or position of the pump relative to the infusion site.
Occlusion Detection Time:
Rate: 16 µL/h Detection Time: 8 h
Rate: 42 µL/h Detection Time: 8 h
The unintended Bolus volume following the release of an Occlusion is ≤40 µL (Typically <15 µL) for all delivery
rates.
Under single fault conditions no more than the following medication is delivered:
10 µL or, if greater,
50% error for deliveries 8 h
25% error for deliveries > 8 h
The system has not been evaluated for use in Oxygen Rich Environments and should not be used in an area
where oxygen is in use.
All parts of the Remunity
Pump for Remodulin
®
(treprostinil) Injection are rated for continuous operation.
The Remunity
Pump for Remodulin
®
(treprostinil) Injection is INTERNALLY POWERED ME
EQUIPMENT.
When the Remunity
Pump for Remodulin
®
(treprostinil) Injection remote interface is being charged it is
CLASS II and INTERNALLY POWERED at all other times.
DKPI-09153-001-Rev. 1.4 213