Pump Idle Attention Alarm
When the pump has
not been delivering
medication for 15
minutes, the Pump
Idle Attention Alarm
displays.
1. Press
to
silence the
attention alarm.
2. Press to
acknowledge the attention alarm.
3. Press and hold the blue pump button until
you hear the four-tone delivering sound.
4. If the pump does not generate the four-tone
delivering sound:
a. Disconnect the infusion set tubing from the
catheter.
b. Disconnect the pump from the cassette.
c. Remove and reinsert the pump battery into
the pump.
d. Listen for the confirmation beep on the
pump.
e. Connect the pump to the cassette.
f. Wait for the self test to complete and then
reconnect the infusion set tubing to the
catheter.
g. Press and hold the blue pump button
until you hear the four-tone sound, or press
on the remote to restart delivery.
Alarms and Attention Alarms
168 DKPI-09153-001-Rev. 1.4