Users Manual Part 7
Problem Possible Causes Corrective Actions
The self test passes, but the pump sounds
an attention alarm tone when the pump
button is pressed.
Delivery rate needs to be set
or updated.
Move the pump or remote
within 12 inches of each
other to enable wireless
communication.
When you reconnect to the
remote, a Basal Not
Started Attention Alarm will
display.
Press
and .
Depending upon the
scenario, you may see a
New Cassette? prompt.
Select Yes or No
accordingly.
Press to start delivery
again.
Table 15: The self test passes, but the pump sounds an Attention Alarm sound when the pump button is pressed
Maintenance and Troubleshooting
192 DKPI-09153-001-Rev. 1.4