Users Manual Part 7
Maintenance and Troubleshooting
Problem Possible Causes Corrective Actions
No fluid is visible at end of infusion set
when priming.
Priming Aid is not fully
rotated.
Fully rotate the arrow on the
priming aid to the lock
symbol on the cassette to
begin priming.
Tubing is occluded Inspect tubing and
straighten any kinked
tubing.
Table 14: No fluid is visible at end of infusion set when priming
DKPI-09153-001-Rev. 1.4 191