Users Manual Part 7

Problem Possible Causes Corrective Actions
Battery Charger status light turns red
when a pump battery is installed.
Pump Battery is too hot or too
cold.
Make sure the temperature
of the battery, or location
where you charge the
battery, is between 59 ºF
(15 ºC) and 95 ºF (35 ºC).
Charging will start
automatically when battery
temperature is within that
range.
Table 9: Battery Charger status light(s) turn red when a pump battery is installed