Users Manual Part 7

Pump Maintenance
Dust, lint, and debris on or in the pump may have
the following effects on the system:
Decrease the water ingress protection rating.
Interfere with the electrical connection between
the pump and battery causing Alarms to occur.
Cause the pump to fail start up tests.
Interfere with the electrical connections
between the remote or pump battery charger
and pump battery causing the battery not to
charge or to charge slowly.
Abrade the surface of the cassette, which can
lead to leaks.
Interfere with rotating locking ring when
connecting or disconnecting the cassette and
pump.
Clean the pump as needed to keep the pump free
of dust, lint, and debris or following exposure to dust
or liquid contaminants.
1. Verify the Dust Cover or a Cassette is attached
to the pump before cleaning.
2. Wet the pump assembly with mild soap and
water and lather it up.
Dish detergent or hand soap is acceptable.
Do not place the pump in a dishwasher.
3. Thoroughly rinse the pump with water.
4. Thoroughly dry the pump with a dry cloth or
paper towel.
Do not rub or press on interior portions of the
pump.
5. Remove the dust cover or cassette and blot
the interior of the pump with dry cloth or paper
towel if any moisture is visible.
Maintenance and Troubleshooting
176 DKPI-09153-001-Rev. 1.4