No Communication
Attention Alarm
When the pump and
remote have not
communicated with
each other for 2
hours, a No
Communication
Attention Alarm
displays with a Verify
pump is delivering
message. This may
be caused by
wireless
communication interference.
1. Press
to silence the attention alarm.
For more information on resolving the
attention alarm, press
.
2. Press to acknowledge the attention alarm.
3. Press and hold the blue pump button until
you hear the four-tone delivering sound:
a. If the pump makes the four-tone delivering
sound, place the pump and remote within
12 inches of each other until communication
is restored. The top of the remote home
screen should change to green.
b. If communication cannot be restored, move
at least 0.8 m (3 ft) from cell phones,
Bluetooth devices, RFID readers, and Wi-Fi
equipment to restore communication.
c. If the pump does not make the four-tone
delivering sound, change the cassette and
pump battery.
See "Cassette Change" on
page59.
Alarms and Attention Alarms
164 DKPI-09153-001-Rev. 1.4