Users Manual Part 6
Depletes Soon Attention
Alarm
When the cassette
will be depleted
within 2 hours, a
Depletes Soon
Attention Alarm
displays.
1. Press
to
silence the
attention alarm.
2. Press to
acknowledge the attention alarm.
3. When the cassette is depleted, change the
cassette and pump battery.
See "Cassette
Change" on page59.
Excessive Noise Attention
Alarm
When your
surrounding
environment is
excessively noisy, an
Excessive Noise
Attention Alarm
displays to alert you
that delivery may be
impacted.
1. Press
to
silence the
attention alarm.
2. Press to acknowledge the attention alarm.
3. To avoid the attention alarm from reappearing,
move to a quieter area.
Alarms and Attention Alarms
162 DKPI-09153-001-Rev. 1.4