Users Manual Part 6
Alarms and Attention Alarms
Delivery Stopped Attention
Alarm
The Delivery
Stopped Attention
Alarm displays to
ensure you are
aware and reminded
that delivery has
been stopped.
This attention alarm
occurs after every
alarm and in
response to the
patient stopping the
pump, such as during the Cassette Change
process.
Make sure to refer to the appropriate alarm,
attention alarm, or cassette change steps to resolve
the delivery stopped attention alarm.
If a delivery stopped attention alarm appears
outside of an alarm, attention alarm, or cassette
change condition follow these steps to restart
delivery.
1. Press
to silence the attention alarm.
2. Press to acknowledge the attention alarm.
3. Press and hold the blue pump button until
you hear the four-tone sound to restart
delivery.
4. If the pump sounds the three-beeps attention
alarm sound,
see "The self test passes, but the
pump sounds an attention alarm tone when the
pump button is pressed." on page192
DKPI-09153-001-Rev. 1.4 161