Users Manual Part 6

Alarms and Attention Alarms
Pump Failure Alarm
When the pump is
broken, a Pump
Failure Alarm
displays. You will
need to switch to
your spare pump
and may need to
request a
replacement from
your specialty
pharmacy.
The top banner
of the screen will flash and the remote will
beep.
1. Press
to silence the alarm.
2. Disconnect infusion set tubing from catheter.
3. Press to acknowledge the alarm.
4. Press to acknowledge the Delivery
Stopped Attention Alarm.
5. Disconnect the pump from cassette.
6. Remove the battery from the pump.
7. Gather your spare pump and insert the pump
battery.
8. Listen for the confirmation beep on the pump.
9. Follow steps to Pair your spare pump with the
remote, reconnect to the cassette, and start
delivery.
See "Pair to a Spare Pump" on
page118.
10. If the spare pump resolves the alarm, contact
your specialty pharmacy to obtain a
replacement pump for the one that failed.
DKPI-09153-001-Rev. 1.4 159