Users Manual Part 6

Pump Error Alarm
When the pump
encounters an error,
a Pump Error Alarm
displays.
The top banner
of the screen
will flash and
the remote will
beep.
1. Press
to
silence the alarm.
2. Disconnect infusion set tubing from catheter.
3. Press to acknowledge the alarm.
4. Press to acknowledge the Delivery
Stopped Attention Alarm.
5. Disconnect the cassette from the pump.
6. Remove and reinsert the pump battery into the
pump.
7. Listen for the confirmation beep on the pump.
8. Disregard the Fill new cassette with X.X mL,
prime, and install message if you want to use
your existing cassette.
a. Reconnect the pump to the used cassette.
b. Wait for the Self Test to complete.
9. Reattach the infusion set tubing to the catheter
and press and hold the blue pump button
or the
on the remote to restart delivery.
10. Press to
select No to the
New Cassette?
question.
11. Press to
start delivery.
Alarms and Attention Alarms
158 DKPI-09153-001-Rev. 1.4