Alarms and Attention Alarms
Cassette Problem Alarm
When your cassette
is broken, a Cassette
Problem Alarm
displays.
The top banner
of the screen
will flash and
the remote will
beep.
1. Press
to
silence the alarm.
2. Disconnect infusion set tubing from catheter.
3. Press to acknowledge the alarm.
4. Press to acknowledge the Delivery
Stopped Attention Alarm.
5. Change the cassette and pump battery
starting from the Cassette Change with
Remote step
"Disconnect Cassette from Pump
and Dispose" on page66.
DKPI-09153-001-Rev. 1.4 153