Users Manual Part 2
Set Up
Enter Delivery Rate
Cell phones, Bluetooth devices, RFID readers,
and Wi-Fi equipment may affect the
communication between the Pump and Remote
or interrupt operation of the Pump. Keep the
Pump and Remote at least 0.8 m [3 ft] away
from these items.
See Table 28 on page223
and See Table 30 on page226 for more
information.
Programming the incorrect delivery rate of
medication can lead to errors in the dose rate
of medication that can lead to harm.
Programming should be performed with the
direction from your healthcare provider.
1. The remote will prompt you to set a delivery
rate (the amount of medication the pump will
deliver per hour of use).
2. Use to set your prescribed rate, then
press .
DKPI-09153-001-Rev. 1.4 57