Users Manual Part 7

Appendix A
Immunity Test IEC 60601 Test Level
Compliance
Level
Electromagnetic Environment -
Guidance
28 V/m
800 MHz 960 MHz
18 Hz PM carrier
28 V/m
1700 MHz 1990 MHz
2400 MHz 2570 MHz
217 Hz PM carrier
9 V/m
5.1 GHz 5.8 GHz
217 Hz PM carrier
28 V/m
800 MHz
960 MHz
18 Hz PM
carrier
28 V/m
1700 MHz
1990 MHz
2400 MHz
2570 MHz
217 Hz PM
carrier
10 V/m
5.1 GHz 5.8
GHz
Cell phones, Wi-Fi equipment,
Bluetooth devices and RFID
readers should be kept at least 0.8
m [3 ft] away from the Remunity
Pump for Remodulin
®
(treprostinil)
Injection.
Table 30: Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
DKPI-09153-001-Rev. 1.4 229